What is the most fixable intramedullary implant for basicervical fracture and transcervical shear fracture? – A finite element study

Open AccessPublished:September 19, 2022DOI:https://doi.org/10.1016/j.jcot.2022.102015

      Abstract

      Objective

      The objectives of this study are 1) to biomechanically compare six different intramedullary fixations for basicervical fracture (AO 31-B3, Type 2 in area classification) and transcervical shear fracture (AO 31-B2.3, Type 1–2 in area classification) using the finite element (FE) method, and 2) to investigate the effects of two different unstable fracture types on fixation.

      Methods

      FE models of two different types of proximal femoral fractures are constructed from CT scan images of a patient with osteoporosis. The fracture models are fixed with a short femoral nail with a single lag screw, short femoral nail with a single blade, and short femoral nail with double lag screws, and then fixed with long femoral nails for each of the three nail types. Subsequently, the maximum loads during walking and stair climbing, as well as the minimum principal strain and compressive failure elements are calculated to assess the fixation of each implant.

      Results

      In both fracture types, the long nail with double lag screws show the smallest volume of compressive failure elements (basicervical fracture, 2 mm3; transcervical shear fracture, 217 mm3). In all types of implants, the volume of the compressive failure elements is larger in the transcervical shear fracture than in the basicervical fracture. A similar trend is observed for the minimum principal strain (compressive strain).

      Conclusion

      The present study shows that a long nail with double lag screws is the most fixative intramedullary nail device for basicervical fracture and transcervical shear fracture in any condition. Furthermore, it is shown that transcervical shear fracture is considerably more unstable than basicervical fracture.

      Keywords

      1. Introduction

      In recent years, the proportion of elderly people in the world's total population as well as the incidence of proximal femoral fractures have increased.
      • Kim S.M.
      • Moon Y.W.
      • Lim S.J.
      • et al.
      Prediction of survival, second fracture, and functional recovery following the first hip fracture surgery in elderly patients.
      ,
      • Dunbar M.J.
      • Howard A.
      • Bogoch E.R.
      • Parvizi J.
      • Kreder H.J.
      Orthopaedics in 2020: predictors of musculoskeletal need.
      Various classification methods for these fractures have been devised to support treatment options for proximal femoral fractures.
      • Muller M.E.
      • Nazarian S.
      • Koch P.
      • Schatzker J.
      The Comprehensive Classification of Fracture of Long Bones.
      • Garden R.S.
      Low-angle fixation in fractures of the femoral neck.
      • Bartonícek J.
      Pauwels' classification of femoral neck fractures: correct interpretation of the original.
      • Evans E.M.
      The treatment of trochanteric fractures of the femur.
      However, because of the wide variety of classification methods, treatment options are diverse and remain non-unified.
      • Schipper I.B.
      • Marti R.K.
      • van der Werken C.
      Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation. Review of literature.
      Therefore, to systematise treatment options, Kijima et al. introduced ‘area classification’, which is a comprehensive classification method for proximal femoral fractures.
      • Kijima H.
      • Yamada S.
      • Konishi N.
      • et al.
      The reliability of classifications of proximal femoral fractures with 3-dimensional computed tomography: the new concept of comprehensive classification.
      In area classification, or in the AO/OTA classification, which is widely used globally, the basicervical fracture (Type 2 in area classification, 31-B3 in AO classification) and transcervical shear fracture (Type 1–2 in area classification, 31-B2.3 in AO classification) exhibit particularly high instability.
      • Kwak D.K.
      • Kim W.H.
      • Lee S.J.
      • Rhyu S.H.
      • Jang C.Y.
      • Yoo J.H.
      Biomechanical comparison of three different intramedullary nails for fixation of unstable basicervical intertrochanteric fractures of the proximal femur.
      ,
      • Li J.
      • Zhao Z.
      • Yin P.
      • Zhang L.
      • Tang P.
      Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis.
      The basicervical fracture is defined as a fracture that occurs medial to the intertrochanteric line above the minor trochanter.
      • Blair B.
      • Koval K.J.
      • Kummer F.
      • Zuckerman J.D.
      Basicervical fractures of the proximal femur. A biomechanical study of 3 internal fixation techniques.
      Moreover, Deneca et al. reported that basicervical fracture have greater rotational instability than that associated with other fractures.
      • Deneka D.A.
      • Simonian P.T.
      • Stankewich C.J.
      • Eckert D.
      • Chapman J.R.
      • Tencer A.F.
      Biomechanical comparison of internal fixation techniques for the treatment of unstable basicervical femoral neck fractures.
      Therefore, the intramedullary nails with double lag screws, including an anti-rotation screw, have been introduced to improve the rotational stability of basicervical fracture.
      • Zelle B.A.
      • Webb A.J.
      • Matson C.
      • et al.
      Safety and efficacy of a two-screw cephalomedullary nail for intertrochanteric femur fracture fixation: a retrospective case series in 264 patients.
      ,
      • Wang Q.
      • Gu X.H.
      • Li X.
      • Wu J.H.
      • Ju Y.F.
      • Huang W.J.
      Management of low-energy basicervical proximal femoral fractures by proximal femoral nail anti-rotation.
      Transcervical shear fractures, also called Pauwels type 3 fractures are prone to shearing force at the angle of the main fracture line. Intramedullary nails reportedly provide greater stability to such fractures than do 3 screws or extramedullary devices.
      • Rupprecht M.
      • Grossterlinden L.
      • Ruecker A.H.
      • et al.
      A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw.
      For these reasons, we mainly treat the Pauwels type 3 fracture with intramedullary nails. Therefore, we investigated what type of intramedullary nails should be used for both types of fractures in this study. Intramedullary nail fixation is a treatment for these types of fractures. The advantages of intramedullary fixation include shorter operative time, less blood loss, and a higher failure load of the implant. Furthermore, it has been reported to be an effective treatment for unstable fractures.
      • Henschel J.
      • Eberle S.
      • Augat P.
      Load distribution between cephalic screws in a dual lag screw trochanteric nail.
      • Zhang Y.
      • He W.
      • Liu Y.W.
      • Feng L.Z.
      Comparison of the effect between eccentric fixation and intramedullary fixation for treatment of intertrochanteric fractures.
      • Weiser L.
      • Ruppel A.A.
      • Nüchtern J.V.
      • et al.
      Extra- vs. intramedullary treatment of pertrochanteric fractures: a biomechanical in vitro study comparing dynamic hip screw and intramedullary nail.
      However, excessive shearing force on the fracture surface causes secondary displacement of bone fragments and, in extreme cases, complications such as ‘cut-out’. In particular, patients with osteoporosis whose central area of the femoral neck often lacks cancellous bone are likely to exhibit decreased fixation, thereby resulting in complications.
      • Hollensteiner M.
      • Sandriesser S.
      • Bliven E.
      • von Rüden C.
      • Augat P.
      Biomechanics of osteoporotic fracture fixation.
      ,
      • Kouvidis G.K.
      • Sommers M.B.
      • Giannoudis P.V.
      • Katonis P.G.
      • Bottlang M.
      Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation.
      Therefore, a strong fixation against shearing force must be ensured at the fracture points, and various types of intramedullary implants are available for this purpose. However, the fixation of intramedullary nails for basicervical and transcervical shear fractures is rarely investigated, and the selection of the optimal intramedullary implant for these fractures remains controversial.
      To determine the most appropriate internal fixation device, fixation must be compared mechanically and quantitatively between devices for the same bone and fracture type. The objectives of this study are 1) to biomechanically compare six different intramedullary fixations for basicervical and transcervical shear fractures using the finite element (FE) method and 2) to investigate the effects of two different unstable fracture types on fixation.

      2. Materials and methods

      This study was approved by the Ethics Committee of the Akita University School of Medicine (Reception No. 2482).

      2.1 FE models

      The subject for the FE models was an osteoporosis patient (age: 73 years, height: 165 cm, weight: 66.2 kg). The patient was diagnosed with osteoporosis because of a proximal femoral fracture due to low-energy trauma. A CT scan (Revolution CT, GE Healthcare, USA) was performed on the subject's lower extremities based on slice thicknesses of 1.25 mm and 512 × 512 pixels per image. In addition, a bone mass phantom (QRM Quality Assurance in Radiology and Medicine GmbH, Baiersdorfer, Germany) was scanned. Table 1 shows the detailed data of bone mineral density obtained using quantitative CT (QCT), which confirmed that the subject was an osteoporotic patient. Mechanical Finder (version 11, Standard Edition) (Research Center of Computational Mechanics, Inc., Tokyo, Japan) was the software to evaluate bone strength, as well as to construct the FE model and perform FE analysis. A three-dimensional (3D) model of the intact femur was constructed by extracting the region of interest around the cortical bone from each CT scan image of the lower extremity that had not fractured.
      Table 1Bone mineral density measured using QCT and literature values
      • Bousson V.D.
      • Adams J.
      • Engelke K.
      • et al.
      In vivo discrimination of hip fracture with quantitative computed tomography: results from the prospective European Femur Fracture Study (EFFECT).
      (mg/cm3).
      Volume of interestPresent studyBousson et al., 2001 (QCT)
      Hip fracture subjects (n = 47), mean (SD)Contrals (n = 60), mean (SD)
      Femoral head199.4182.2 (44.7)237.1 (52.3)
      Femoral neck260.1242.5 (48.7)291.5 (48.3)
      Next, based on area classification,
      • Kijima H.
      • Yamada S.
      • Konishi N.
      • et al.
      The reliability of classifications of proximal femoral fractures with 3-dimensional computed tomography: the new concept of comprehensive classification.
      the fracture line of the basicervical fracture (Type 2 in area classification, 31-B3 in AO classification) and transcervical shear fracture (Type 1–2 in area classification, 31-B2.3 in AO classification) were applied to the 3D femur model (Fig. 1). To fix these fractures, six types of intramedullary implants were used. We classified the intramedullary nail fixation modes into three groups based on the device to be inserted into the femoral head: Trochanteric Fixation Nail Advanced (TFNA) with a single blade (DePuy Synthes, Warsaw, IN, USA) (Group 1), Gamma 3 nail with a single lag screw (Stryker, Mahwah, NJ, USA) (Group 2), and Cephalomedullary Asia nail with double lag screws (Zimmer Biomet, Warsaw, IN, USA) (Group 3); short and long nails were used for each group (Fig. 2). The 3D models of each nail were constructed using the same method as the bone model, based on CT images captured from the nail. Because it was difficult to reproduce the fine regions of the shape from the CT images, the screw, blade, and distal locking screw were created using 3D CAD software Fusion360 (Autodesk, Inc., San Rafael, USA). The sizes of the implants are shown in Table 2. Four-node tetrahedral solid elements with minimum and maximum element sizes of 1 and 8 mm, respectively, was used to mesh the inner side of the cancellous and cortical bone. In addition, shell elements with a thickness of 0.3 mm were used on the outer surface of the cortical bone.
      • Bessho M.
      • Ohnishi I.
      • Okazaki H.
      • Sato W.
      • Kominami H.
      • Matsunaga S.
      Prediction of the strength and fracture location of the femoral neck by CT-based finite-element method: a preliminary study on patients with hip fracture.
      To ensure the reliability of the analytical results, a mesh convergence test was performed. Four different models were constructed for each of the two fracture types, with minimum mesh sizes of 0.9, 1, 2, and 3 mm, respectively. The displacement of the femoral head was evaluated when a vertical load of 1500 N was applied, with an increase of <5% as the criterion for convergence. For both basicervical fracture and transcervical shear fracture, the convergence criteria were fulfilled between 2 and 1 mm. However, because of the difficulty in shaping the screw threads at 2 mm, a minimum mesh size of 1 mm was selected. The average number of nodes, shell elements, and solid elements in the FE models were 160204, 73726, and 772690, respectively.
      Fig. 1
      Fig. 1Positions of main fracture line: (a) Basicervical fracture (AO 31-B3, Type 2 in area classification); (b) transcervical shear fracture (AO 31-B2.3, Type1-2 in area classification).
      Fig. 2
      Fig. 23D model with implants inserted in Type 2: (a) Short nail of Group 1; (b) long nail of Group 1; (c) short nail of Group 2; (d) long nail of Group 2; (e) short nail of Group 3; (f) long nail of Group 3. Group 1: one blade; Group 2: one lag screw; Group 3: double lag screws.
      Table 2Size of implants.
      ComponentsGroup 1Group 2Group 3
      Short nailLong nailShort nailLong nailShort nailLong nail
      Lag screwLength (mm)909090909090
      Diameter (mm)101010.510.511, 511, 5
      NailLength (mm)170260180215170235
      Diameter (mm)121211.511.51111
      Distal locking screwLength (mm)404040404040
      Diameter (mm)444444

      2.2 Material properties

      To simulate the subject's specific bone structure, the BMD of each element was determined using Hounsfield unit (HU) values of the CT images (Fig. 3). The relationship between the HU and BMD values was calibrated using a bone mass phantom imaged with the subject's lower extremities to accurately reflect the BMD values in the model. The equation that describes the relationship between the HU value and BMD, i.e., ρ (mg/cm3), is as follows:
      ρ={0.0(HU<1)0.88×HU17.1(HU1)
      (1)


      Fig. 3
      Fig. 3Distribution of (a) bone mineral density and (b) elastic modulus in femur cross-section.
      To determine the apparent Young's modulus E (MPa) of each element of the bone, we used the conversion formula proposed by Keyak et al.
      • Keyak J.H.
      • Rossi S.A.
      • Jones K.A.
      • Skinner H.B.
      Prediction of femoral fracture load using automated finite element modeling.
      as follows:
      E={0.001(ρ=0)33900ρ2.20(0<ρ0.27)5370ρ+469(0.27<ρ<0.6)10200ρ2.01(0.6ρ)
      (2)


      The Poisson's ratio of the bone was set to 0.4, and each element was assigned homogeneity.
      • Kubo Y.
      • Motomura G.
      • Utsunomiya T.
      • et al.
      Distribution of femoral head subchondral fracture site relates to contact pressures, age, and acetabular structure.
      All implants were assumed to be made of titanium alloy (Ti–6Al–4V), where a Young's modulus of 113.8 MPa,
      • Li J.
      • Han L.
      • Zhang H.
      • et al.
      Medial sustainable nail versus proximal femoral nail antirotation in treating AO/OTA 31-A2.3 fractures: finite element analysis and biomechanical evaluation.
      a Poisson's ratio of 0.3425 were assigned to each element homogeneously.

      2.3 Boundary conditions

      The loading conditions are presented in Table 3. The maximum loads during walking and stair climbing calculated by Heller et al.
      • Heller M.O.
      • Bergmann G.
      • Kassi J.P.
      • Claes L.
      • Haas N.P.
      • Duda G.N.
      Determination of muscle loading at the hip joint for use in pre-clinical testing.
      were applied to each FE model, as shown in Fig. 4. The load profile included the muscle forces at the femur, normalized by the body weight. Each of the values of the load profile was calculated using the musculoskeletal model and validated using in vivo data. To prevent the displacement of the rigid body, the distal end of the femur was fully constrained with six degrees of freedom. The coefficient of friction was 0.3 between the bone and implant,
      • Zeng W.
      • Liu Y.
      • Hou X.
      Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis.
      ,
      • Chen J.
      • Ma J.X.
      • Wang Y.
      • et al.
      Finite element analysis of two cephalomedullary nails in treatment of elderly reverse obliquity intertrochanteric fractures: zimmer natural nail and proximal femoral nail antirotation-ΙΙ.
      and 0.46 between bones.
      • Zeng W.
      • Liu Y.
      • Hou X.
      Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis.
      ,
      • Chen J.
      • Ma J.X.
      • Wang Y.
      • et al.
      Finite element analysis of two cephalomedullary nails in treatment of elderly reverse obliquity intertrochanteric fractures: zimmer natural nail and proximal femoral nail antirotation-ΙΙ.
      Meanwhile, between-implant components were set as tie conditions (bonded to each other).
      Table 3Muscular forces on proximal femur based on walking and stair climbing.
      • Bousson V.D.
      • Adams J.
      • Engelke K.
      • et al.
      In vivo discrimination of hip fracture with quantitative computed tomography: results from the prospective European Femur Fracture Study (EFFECT).
      ,
      • Bessho M.
      • Ohnishi I.
      • Okazaki H.
      • Sato W.
      • Kominami H.
      • Matsunaga S.
      Prediction of the strength and fracture location of the femoral neck by CT-based finite-element method: a preliminary study on patients with hip fracture.
      ForcePositionWalkingStair climbing
      Fx (N)Fy (N)Fz (N)Fx (N)Fy (N)Fz (N)
      Body weightP100−64900−649
      Hip contactP1350213−1488386393−1534
      Internal resultantP15383−50884182−455
      AbductorP2−376−28561−455−187551
      Ilio-tibial tract, proximal partP2−68−1983
      Ilio-tibial tract, distal partP235−109
      Tensor fascia latae, proximal partP2−47−7586−20−3219
      Tensor fascia latae, distal partP235−12312−42
      Vastus lateralisP36−120−60314−145−877
      Vastus medialisP457−257−1733

      2.4 Nonlinear fracture analysis

      Nonlinear FE analysis was performed to include the yield failure of the trabecular bone elements due to compression. The yield strength in compression was based on the yield criterion proposed by Keyak et al.
      • Keyak J.H.
      • Lee I.Y.
      • Skinner H.B.
      Correlations between orthogonal mechanical properties and density of trabecular bone: use of different densitometric measures.
      The modulus of elasticity after yielding was set to 5% of E.
      • Wakao N.
      • Harada A.
      • Matsui Y.
      • et al.
      The effect of impact direction on the fracture load of osteoporotic proximal femurs.
      Fracture of the element in compression was defined as the case in which the minimum principal strain of the element was less than −10000.
      • Wakao N.
      • Harada A.
      • Matsui Y.
      • et al.
      The effect of impact direction on the fracture load of osteoporotic proximal femurs.

      2.5 Selection of evaluation criteria

      In this study, the minimum principal strain (compressive strain) and compressive failure element at the fractured fragment (femoral head) were calculated to assess the stability of the fixed fractures. The peri-implant element compressive fracture in the bone fragment (femoral head) was evaluated by assuming a larger volume of the compressive fracture elements.
      • Goffin J.M.
      • Pankaj P.
      • Simpson A.H.
      The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study.
      ,
      • Goffin J.M.
      • Pankaj P.
      • Simpson A.H.
      A computational study on the effect of fracture intrusion distance in three- and four-part trochanteric fractures treated with Gamma nail and sliding hip screw.
      The larger the volume of the compressive fracture elements, the greater is the risk of displacement of the fractured fragment.

      3. Results

      3.1 Minimum principal strain (compressive strain) of head and neck regions

      Fig. 5 shows the distribution of the minimum principal strains in the head and neck regions. The bone fragments showed high compressive strain (i.e. lower minimum principal strain) near the screw tip in basicervical fracture and above the screw in transcervical shear fracture. Meanwhile, on the femur side, high compressive strain occurred in the lower region of the screw. It was revealed that the strain site and strain volume differed significantly, although the fracture lines were only slightly different.
      Fig. 5
      Fig. 5Minimum principal strain distributions of head and neck regions: (a) Short nail of Group 1; (b) long nail of Group 1; (c) short nail of Group 2; (d) long nail of Group 2; (e) short nail of Group 3; (f) long nail of Group 3.

      3.2 Minimum principal (compressive) strain of fracture surface

      Fig. 6 shows the distribution of the minimum principal strain on the fracture surface. In both the basicervical fracture and transcervical shear fracture, the highly compressive strain (i.e. lower minimum principal strain) tended to be concentrated superior to the screw because of the shearing force at the fracture regions. Fig. 8 shows the minimum value of the minimum principal strain (i.e. maximum compressive strain) on the fracture surface. The short nails showed lower values than the long nails for both basicervical and transcervical shear fractures. Comparing the average values of walking and stair climbing, basicervical fracture showed the lowest value (−0.018) for Group 2 with a short nail and the highest value (−0.012) for Group 3 with a long nail. Conversely, transcervical shear fracture showed the lowest value (−0.288) for Group 1 with a short nail and the highest value (−0.162) for Group 3 with a long nail.
      Fig. 6
      Fig. 6Minimum principal strain distributions on fracture surfaces of fragment (femoral head): (a) Short nail of Group 1; (b) long nail of Group 1; (c) short nail of Group 2; (d) long nail of Group 2; (e) short nail of Group 3; (f) long nail of Group 3.

      3.3 Compressive fracture element at fragment

      The region where compressive failure occurred was the same as the region where a high compressive strain occurred (Fig. 5, Fig. 7). The total volume of the compressive failure elements was larger for stair climbing than for walking (Fig. 9). In basicervical fracture, the minimum volume of the compressive failure elements was recorded in Group 3 with a long nail (2 mm3), where the maximum was recorded in Group 1 with a long nail (35 mm3). Meanwhile, in transcervical shear fracture, the minimum was recorded in Group 3 with a long nail (217 mm3), whereas the maximum was recorded in Group 2 with a short nail (1169 mm3).
      Fig. 7
      Fig. 7Distribution of compressive failure elements: (a) Short nail of Group 1; (b) long nail of Group 1; (c) short nail of Group 2; (d) long nail of Group 2; (e) short nail of Group 3; (f) long nail of Group 3.
      Fig. 8
      Fig. 8Minimum value of minimum principal strain at each fracture surface: (A) Basicervical fracture; (B) transcervical shear fracture.
      Fig. 9
      Fig. 9Volume of compressive failure elements in (A) Basicervical fracture and (B) transcervical shear fracture.

      4. Discussion

      Low BMD is a significant factor that increases the risk of failure after surgery.
      • Hedström M.
      Are patients with a nonunion after a femoral neck fracture more osteoporotic than others? BMD measurement before the choice of treatment?: a pilot study of hip BMD and biochemical bone markers in patients with femoral neck fractures.
      To achieve better fixation for osteoporotic proximal femoral fractures, an optimal device must be selected.
      • Wang Q.
      • Gu X.H.
      • Li X.
      • Wu J.H.
      • Ju Y.F.
      • Huang W.J.
      Management of low-energy basicervical proximal femoral fractures by proximal femoral nail anti-rotation.
      In this study, we compared the fixation of six different intramedullary implants for two types of fractures (basicervical and transcervical shear fractures) using an osteoporotic bone model. Our findings indicated that long nails with double lag screws afforded the best fixation for both fractures. Furthermore, the long nail afforded better fixation against shearing forces in the fracture than the short nail. The results of this study are valuable for the selection of treatments for two types of unstable fractures.
      It was discovered that the compressive strain at the fracture site was advantageous to the bone healing process; however, Peter et al.
      • Augat P.
      • Burger J.
      • Schorlemmer S.
      • Henke T.
      • Peraus M.
      • Claes L.
      Shear movement at the fracture site delays healing in a diaphyseal fracture model.
      reported that the shearing force at the fracture site was a disadvantage. In the present analysis, a high compressive strain occurred superior to the screw in the femoral head, and a high compressive strain occurred inferior to the screw on the femoral side (Fig. 5). This indicates that a high compressive strain was generated around the screw owing to the shearing force in the fracture region (Fig. 6). Furthermore, excessive compressive strain resulted in the microfracture of the trabecular bone, which resulted in complications such as secondary displacement and cut-out.
      • Goffin J.M.
      • Pankaj P.
      • Simpson A.H.
      The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study.
      ,
      • Hedström M.
      Are patients with a nonunion after a femoral neck fracture more osteoporotic than others? BMD measurement before the choice of treatment?: a pilot study of hip BMD and biochemical bone markers in patients with femoral neck fractures.
      This suggests that the compressive strain in the fracture surface represents the fixation against the force in the shear direction of the fracture region.
      Kwak et al. reported that fixation achieved using a blade and screw yielded better outcomes as compared with fixation using a single blade for basicervical fracture.
      • Wang Q.
      • Gu X.H.
      • Li X.
      • Wu J.H.
      • Ju Y.F.
      • Huang W.J.
      Management of low-energy basicervical proximal femoral fractures by proximal femoral nail anti-rotation.
      Our results showed that better fixation was achieved using double lag screws compared with using single lag screws for both basicervical and transcervical shear fractures (Fig. 3, Fig. 4). In both fractures, shearing forces were generated at the fracture surface when a load was applied. Therefore, it is suggested that the compressive force generated around the screw is more dispersed in two devices than in the case of one device that pierces the femoral head, i.e. the implant with a single lag screw or single blade. This tendency was confirmed by the fact that the difference in the volume of compressive fracture between the loading conditions was smaller when two devices were inserted into the femoral head than when one device was inserted.
      Previous studies reported that better fixation was obtained by increasing the intramedullary nail length in various fracture types.
      • Hong J.B.
      • Dan Y.
      • Ouyang L.
      • et al.
      Biomechanical study on different lengths of PFNA fixation for unstable intertrochanteric femoral fractures.
      ,
      • Liang C.
      • Peng R.
      • Jiang N.
      • Xie G.
      • Wang L.
      • Yu B.
      Intertrochanteric fracture: association between the coronal position of the lag screw and stress distribution.
      In also basicervical and transcervical shear fractures, which are known to be highly unstable, long nails offered better fixation than short nails, and this tendency is more evident in transcervical shear fractures. The distribution of the minimum principal strain showed that the red area of the large compressive strain (Fig. 3) was concentrated in the upper region of the lag screw, and this tendency was more evident in the transcervical shear fracture. Generally, femoral neck fractures are more unstable as the main fracture line becomes vertical.
      • Wang G.
      • Tang Y.
      • Wu X.
      • Yang H.
      Finite element analysis of a new plate for Pauwels type III femoral neck fractures.
      Our results show that transcervical shear fracture, whose main fracture line is more vertical, is more unstable than basicervical fracture. This finding supports common understanding. In addition, comparing the volume of compressive failure elements in each implant, the maximum value was approximately 22 and 5 times higher than the minimum values in the basicervical fracture and transcervical shear fracture, respectively. This indicates the importance of selecting an optimal device.
      This study has several limitations. First, the number of participants was small. Further studies should be conducted to investigate the effects of changes in bone shape and quality. Second, the contact area between the implant components was fixed. However, in reality, because a sliding mechanism exists in the contact area, friction in the contact area should be defined and analysed. Third, only the main fracture line was applied to each finite element model. In fact, bone fragments were generated at the fracture site.
      • Furui A.
      • Terada N.
      • Mito K.
      Mechanical simulation study of postoperative displacement of trochanteric fractures using the finite element method.
      The reproduction of fractures will be investigated in future studies.

      5. Conclusions

      The present study showed that a long intramedullary nail with double lag screws was the most effective implant for the fixation of proximal femoral basicervical fracture (AO 31-B3, Type 2 in area classification) and transcervical shear fracture (AO 31-B2.3, Type1-2 in area classification). In addition, it was proven for the first time that transcervical shear fracture is a significantly more unstable fracture than basicervical fracture.

      Declarations of competing interest

      None.

      References

        • Kim S.M.
        • Moon Y.W.
        • Lim S.J.
        • et al.
        Prediction of survival, second fracture, and functional recovery following the first hip fracture surgery in elderly patients.
        Bone. 2012; 50: 1343-1350
        • Dunbar M.J.
        • Howard A.
        • Bogoch E.R.
        • Parvizi J.
        • Kreder H.J.
        Orthopaedics in 2020: predictors of musculoskeletal need.
        J Bone Joint Surg Am. 2009; 91: 2276-2286
        • Muller M.E.
        • Nazarian S.
        • Koch P.
        • Schatzker J.
        The Comprehensive Classification of Fracture of Long Bones.
        Springer–Verlag, Berlin1990 (145–145)
        • Garden R.S.
        Low-angle fixation in fractures of the femoral neck.
        J Bone Joint Surg Br. 1961; 43: 647-663
        • Bartonícek J.
        Pauwels' classification of femoral neck fractures: correct interpretation of the original.
        J Orthop Trauma. 2001; 15: 358-360
        • Evans E.M.
        The treatment of trochanteric fractures of the femur.
        J Bone Joint Surg Br. 1949; 31: 190-203
        • Schipper I.B.
        • Marti R.K.
        • van der Werken C.
        Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation. Review of literature.
        Injury. 2004; 35: 142-151
        • Kijima H.
        • Yamada S.
        • Konishi N.
        • et al.
        The reliability of classifications of proximal femoral fractures with 3-dimensional computed tomography: the new concept of comprehensive classification.
        Adv Orthop. 2014; 359689
        • Kwak D.K.
        • Kim W.H.
        • Lee S.J.
        • Rhyu S.H.
        • Jang C.Y.
        • Yoo J.H.
        Biomechanical comparison of three different intramedullary nails for fixation of unstable basicervical intertrochanteric fractures of the proximal femur.
        Experimental Studies. 2018; 7618079
        • Li J.
        • Zhao Z.
        • Yin P.
        • Zhang L.
        • Tang P.
        Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis.
        J Orthop Surg Res. 2019; 14: 1-8
        • Blair B.
        • Koval K.J.
        • Kummer F.
        • Zuckerman J.D.
        Basicervical fractures of the proximal femur. A biomechanical study of 3 internal fixation techniques.
        Clin Orthop Relat Res. 1994; 306: 256-263
        • Deneka D.A.
        • Simonian P.T.
        • Stankewich C.J.
        • Eckert D.
        • Chapman J.R.
        • Tencer A.F.
        Biomechanical comparison of internal fixation techniques for the treatment of unstable basicervical femoral neck fractures.
        J Orthop Trauma. 1997; 11: 337-343
        • Zelle B.A.
        • Webb A.J.
        • Matson C.
        • et al.
        Safety and efficacy of a two-screw cephalomedullary nail for intertrochanteric femur fracture fixation: a retrospective case series in 264 patients.
        Patient Saf Surg. 2018; 12: 31
        • Wang Q.
        • Gu X.H.
        • Li X.
        • Wu J.H.
        • Ju Y.F.
        • Huang W.J.
        Management of low-energy basicervical proximal femoral fractures by proximal femoral nail anti-rotation.
        Orthop Surg. 2019; 11: 1173-1179
        • Rupprecht M.
        • Grossterlinden L.
        • Ruecker A.H.
        • et al.
        A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw.
        J Trauma. 2011; 71: 625-634
        • Henschel J.
        • Eberle S.
        • Augat P.
        Load distribution between cephalic screws in a dual lag screw trochanteric nail.
        J Orthop Surg Res. 2016; 11: 1-10
        • Zhang Y.
        • He W.
        • Liu Y.W.
        • Feng L.Z.
        Comparison of the effect between eccentric fixation and intramedullary fixation for treatment of intertrochanteric fractures.
        Zhong Guo Gu Shang. 2015; 28: 117-121
        • Weiser L.
        • Ruppel A.A.
        • Nüchtern J.V.
        • et al.
        Extra- vs. intramedullary treatment of pertrochanteric fractures: a biomechanical in vitro study comparing dynamic hip screw and intramedullary nail.
        Arch Orthop Trauma Surg. 2015; 135: 1101-1106
        • Hollensteiner M.
        • Sandriesser S.
        • Bliven E.
        • von Rüden C.
        • Augat P.
        Biomechanics of osteoporotic fracture fixation.
        Curr Osteoporos Rep. 2019; 17: 363-374
        • Kouvidis G.K.
        • Sommers M.B.
        • Giannoudis P.V.
        • Katonis P.G.
        • Bottlang M.
        Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation.
        J Orthop Surg Res. 2009; 4: 1-9
        • Bousson V.D.
        • Adams J.
        • Engelke K.
        • et al.
        In vivo discrimination of hip fracture with quantitative computed tomography: results from the prospective European Femur Fracture Study (EFFECT).
        J Bone Miner Res. 2011; 26: 881-893
        • Bessho M.
        • Ohnishi I.
        • Okazaki H.
        • Sato W.
        • Kominami H.
        • Matsunaga S.
        Prediction of the strength and fracture location of the femoral neck by CT-based finite-element method: a preliminary study on patients with hip fracture.
        J Orthop Sci. 2004; 9: 545-550
        • Keyak J.H.
        • Rossi S.A.
        • Jones K.A.
        • Skinner H.B.
        Prediction of femoral fracture load using automated finite element modeling.
        J Biomech. 1998; 31: 125-133
        • Kubo Y.
        • Motomura G.
        • Utsunomiya T.
        • et al.
        Distribution of femoral head subchondral fracture site relates to contact pressures, age, and acetabular structure.
        AJR Am J Roentgenol. 2020; 215: 448-457
        • Li J.
        • Han L.
        • Zhang H.
        • et al.
        Medial sustainable nail versus proximal femoral nail antirotation in treating AO/OTA 31-A2.3 fractures: finite element analysis and biomechanical evaluation.
        Injury. 2019; 50: 648-656
        • Heller M.O.
        • Bergmann G.
        • Kassi J.P.
        • Claes L.
        • Haas N.P.
        • Duda G.N.
        Determination of muscle loading at the hip joint for use in pre-clinical testing.
        J Biomech. 2005; 38: 1155-1163
        • Zeng W.
        • Liu Y.
        • Hou X.
        Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis.
        Comput Methods Progr Biomed. 2020; 196105714
        • Chen J.
        • Ma J.X.
        • Wang Y.
        • et al.
        Finite element analysis of two cephalomedullary nails in treatment of elderly reverse obliquity intertrochanteric fractures: zimmer natural nail and proximal femoral nail antirotation-ΙΙ.
        J Orthop Surg Res. 2020; 14: 1-8
        • Keyak J.H.
        • Lee I.Y.
        • Skinner H.B.
        Correlations between orthogonal mechanical properties and density of trabecular bone: use of different densitometric measures.
        J Biomed Mater Res. 1994; 28: 1329-1336
        • Wakao N.
        • Harada A.
        • Matsui Y.
        • et al.
        The effect of impact direction on the fracture load of osteoporotic proximal femurs.
        Med Eng Phys. 2009; 31: 1134-1139
        • Goffin J.M.
        • Pankaj P.
        • Simpson A.H.
        The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study.
        J Orthop Res. 2013; 31: 596-600
        • Goffin J.M.
        • Pankaj P.
        • Simpson A.H.
        A computational study on the effect of fracture intrusion distance in three- and four-part trochanteric fractures treated with Gamma nail and sliding hip screw.
        J Orthop Res. 2014; 32: 39-45
        • Hedström M.
        Are patients with a nonunion after a femoral neck fracture more osteoporotic than others? BMD measurement before the choice of treatment?: a pilot study of hip BMD and biochemical bone markers in patients with femoral neck fractures.
        Acta Orthop Scand. 2004; 75: 50-52
        • Augat P.
        • Burger J.
        • Schorlemmer S.
        • Henke T.
        • Peraus M.
        • Claes L.
        Shear movement at the fracture site delays healing in a diaphyseal fracture model.
        J Orthop Res. 2003; 21: 1011-1017
        • Hong J.B.
        • Dan Y.
        • Ouyang L.
        • et al.
        Biomechanical study on different lengths of PFNA fixation for unstable intertrochanteric femoral fractures.
        J Musculoskelet Neuronal Interact. 2017; 17: 299-302
        • Liang C.
        • Peng R.
        • Jiang N.
        • Xie G.
        • Wang L.
        • Yu B.
        Intertrochanteric fracture: association between the coronal position of the lag screw and stress distribution.
        Asian J Surg. 2018; 41: 241-249
        • Wang G.
        • Tang Y.
        • Wu X.
        • Yang H.
        Finite element analysis of a new plate for Pauwels type III femoral neck fractures.
        J Int Med Res. 2020; 48300060520903669
        • Furui A.
        • Terada N.
        • Mito K.
        Mechanical simulation study of postoperative displacement of trochanteric fractures using the finite element method.
        J Orthop Surg Res. 2018; 13: 1-8